Combining Function and Position

 

The Positron Emission Tomography/Computed Tomography Scanner

 

The medical Imaging Device of the Future?

 

By Jack Wells

 

 

 

 

 

 

Abstract

 

Separate PET and CT scanners are widely used in the clinical environment to produce functional and anatomical images respectively. The dual scanner is a recent innovation and combines both scans into one examination. Information from both scans can be aligned more easily onto one image.  The device has led to significant improvements in patient administration. Most importantly it has improved the accuracy of tumour detection and localisation. Studies revealed that the hybrid scanner aided tumour identification in 56% of lung cancer patients. In addition it was discovered that the increased accuracy of the device led to changes in radiation treatment planning in 40% of the patients sampled. However the accuracy of the images is limited due to the finite movements of the patient during the CT scan.

 

 

 

 

 

 

 

 

 

 

 

 

The PET/CT scanner is a tumour detecting imaging system that combines anatomical and functional information to produce a comprehensive picture of the internal workings of the body.

Liver cancer

    CT image             PET image      Combined Image

Figure 1: Metastasis of the liver and pelvis (1)

 

As illustrated by figure 1, this technique can produce images with greater clarity then either method alone. This is of obvious importance to the detection of tumors and significantly influences the effectiveness of hospital treatment. Furthermore, the scanners have additional advantages that benefit the clinical environment.

 

I will begin by explaining the principles behind PET and CT. The second section will discuss image registration, how the information from each device is combined. Next, I will detail the design and features of the scanners. Finally, the benefits and disadvantages of the system are reviewed and the current state of the field is summarized.

 

Computer Tomography (CT) Imaging

 

The principle of CT relies on synchronised movement of the source and detector to produce an image in the plane of the patient. The word tomography originates from the Greek τομοs (slice) an dγράφειυ (to write) [20]. A set of thin x ray lines is used to scan the required area to be imaged.  The process is then repeated for a large number of angles. This results in attenuation readings for all possible angles and distances from the required plane. These results can then yield to the reconstruction of the actual attenuation at each point. 

 

 

Figure 2 [21]: A schematic illustration of the progressive generations of CT geometries. Each set of lines from the source to detector represents a view. The axis of rotation is always the centre of the subject.

 

A. First-generation, translate-rotate pencil beam geometry. B. Second-generation, translate-rotate fan beam geometry. C. Third-generation, rotate-only geometry. D. Third-generation offset-mode geometry. 

 

 

The standard protocol for the dual scanner is a low dose spiral CT [24]. The X-ray tube rotates around the patient. The patient is then moved through the central axis of the rotation. Therefore in the frame of the patient, the tube possesses a helical orbit.                                                              

 

X-ray source

 

X-rays are produced when energetic electrons are stopped in a material with a high atomic number [22]. Electrons are emitted from a heated tungsten filament. These are accelerated via a potential difference in a vacuum to a tungsten target (molybdenum is used to produce a lower energy spectrum for mammography). The source must provide a very short, high energy burst of x-rays so the image will not be blurred by patient movement. In addition the size of the source must be small to ensure the edges of the image appear sharp [22]. This must be achieved despite the fact that 95%-99% of the electron energy is transferred to heat within the target. Therefore a rotating anode is used to dissipate the heat over a wider area. In addition the target is inclined to ensure a small focal spot whilst maximising the area available to the incident electrons. The x-ray spectrum produced is made up of characteristic and bremstrahlung (continuous) radiation. The maximum energy of the x-rays is proportional to the accelerating voltage and the intensity is proportional to the current and is reliant on the voltage squared. The target material and filtration due to the surrounding casing further shapes the outgoing spectrum.  

 

The focal spot, energy spectrum and intensity are the main variables which must be considered for an effective x-ray source. The focal spot determines the spatial resolution of the image as overlapping beams will result in blurring of the views. The energy of the incident photons establishes their penetrative ability. X-rays with greater frequencies are more likely to be detected but are less able to distinguish between different tissue types. Increasing the intensity enhances the signal to noise ratio at the detector but raises the dose to the patient. Furthermore greater intensities require a larger focal spot. It is clear that the design of the source must account for a compromise of these variables to produce optimal performance. Modern CT scanners have focal spots of 0.5mm to 2mm.

 

Interactions of x-rays within the patient and detector [22]

 

At low photon energies within high atomic number materials, the photoelectric effect predominates. The Compton Effect dominates as the incident energy increases in all materials. The reliance of the interactions upon the physical properties of the tissue and detector is the essential aspect of CT imaging that allows images to be produced. The probability of a photoelectric interaction taking place is proportional to the atomic number of the material to the power of four.  The probability of Compton interactions is proportional to the electron density and therefore the physical density of the medium. Consequently, low energy x-rays are better at distinguishing between different tissues as the photoelectric effect predominates. However this does increase the dose to the patient as less x-rays penetrate through to the detector.  The combination of both these effects gives the total attenuation. 

 

Conventional X-ray systems tend to use a simple film/screen combination. However, CT detectors are made up of gas proportional chambers or rows of scintillator and photomultiplier pairs. The operation of the latter is described in the next section. This allows information to be stored digitally which is essential for the superposition of images.  In 1994 George Charpak received the Nobel Prize for physics for the multi wire proportional chamber (MWPC). A Russian group based in Novosibrsk, later collaborated with George to develop a medical imaging system based on the MWPC.  The most unique aspect of such a devise is that it detects each X-ray photon as a single event rather then the superposition of a large number of interactions. As a results x-ray photons can be clearly distinguished from noise. Thus the signal to noise ratio is reduced and a high quality image is produced with reduced dose to the patient.  The chambers contain pressurised Xenon gas. Within the chambers are a series of wires with a high positive potential. The outside of the chamber is negatively charged. The wires are roughly 1mm apart. Incident X-ray photons will interact with the Xenon gas and create charge pairs. Under the influence of the electric field these accelerate towards the wires. Consequently they undergo multiple collisions creating more charge pairs and thus produce a charge cloud. This interacts with the wire to produce a pulse which is counted if its magnitude is sufficient. A computer records the number of counts on each wire every 30ms. Gas ionisation chambers have high stability and inherent collimation. However the majority of CT devises relay on scintillator/ photomultiplier pairs as they can be made into large arrays and a re relatively cheap.

 

As a consequence of scattering effects, the greatest test of x-ray imaging is the detection of low contrast details. Consequently tissue types with linear attenuation coefficients differing by less then 5% are often indistinguishable. This is where the PET scanner is invaluable as physicians attempt to identify pathological effects.  

 

 

 

Positron Emission Tomography (PET)

 

  Figure 2 : A schematic diagram of a PET scanner (3)

 

 

The main attribute of PET imaging is its capability to trace radioactive labels metabolised in the tissue to provide information about its biochemical and physiological behaviour (4).  The most common positron emitter is 18F. This is administered as a radiopharmaceutical, named fluorodeoxyglucose (FDG). Cancer cells tend to accumulate more glucose in comparison to other tissues. Consequently, there will be higher concentrations of FDG in tumours. The radioactive nuclei then decay by positron emission [25]. A nuclear proton decays into a positron and a neutron. The atomic mass of the atom does not change but its atomic number decreases by one.  An ejected positron then annihilates with an electron. This produces photons which travel in opposite directions. Each photon has energy of 511kev, equal to the rest mass of the respective ferminions.  The finite time window used in coincidence detection involving two surrounding detectors placed opposite to each other provides an efficient mechanism of electronic collimation (3). Scattered photons do not tend to arrive within the coincidence time window and thus are rejected. Once 100,000 or more events are recorded, the distribution of the tracer is computed using tomographic reconstruction techniques [25].

 

A scintillation crystal converts the gamma rays to visible light photons. On average, it requires 30eV of light to produce a visible light photon in NaI (the most commonly used scintillation material) [26]. Therefore the crystal is around 10% efficient. The precise energy required varies between resulting in a spread in the number of visible photons produced for a given energy absorbed. However this uncertainty is largely negated by the statistical ambiguity in the photomultiplier tube. 

 

The photomultiplier tube converts an incoming signal of a few hundred or thousand visible light photons to an electrical signal. It consists of the photocathode and the multiplier. The photocathode is designed to produce a large number of low energy electrons whilst ensuring the outgoing current is proportional to the intensity of the signal from the crystal. An electron absorbs a visible photon. The electron then travels to the surface of the photocathode. Finally the electron escapes from the surface. These electrons are then electrically focused to collide with the first electrode of the multiplier. The acceleration of the electrons by a potential difference causes secondary electronic emission at the electrode. The multiplication factor at each junction is between 5 and 10 [26]. For each electrode there is an optimum accelerating voltage. The greater the voltage, the more secondary electrons are produced. However they originate at greater depths within the material and thus are less likely to escape. The gain is independent of the number of electrons from the photocathode from 1 up to a few thousand. Thus a linear output can be obtained from an increasing number of light photons [26].      

 

The recent progression of new and smaller crystals, such as LSO, GSO (with better timing and energy properties), has increased the accuracy of the images (5). At this time the major clinical application of PET is tumour identification and localisation in oncology.

 

 

Image Registration

 

Data from both devices must be correctly combined to produce accurate images. The main condition for registering the information is that a relationship between the coordinates of the corresponding points in the two images is determined (1). This is found using a coordinate transfer function (CTF), which is a set of operators which map pixel values from one image to the other. This is illustrated by diagram 1:

 

 

The poor resolution of PET images means that they often contain few anatomical landmarks. Previously it was necessary to view them in conjunction with CT images with the patient often returning on a separate day. Reference markers, attached to the surface of the patient, were used to account for the differences in subject position. This was suitable for fixed organs such as the brain but other sections such as the abdomen are mobile and may shift position between data acquisition.  

 

The introduction of the dual scanner has simplified image registration. The two sets of data are collected in sequence and so are intrinsically registered. As a result only the distance between the CT and PET sources and detectors must be considered.

 

The design and operation of the scanner

 

Prior to the introduction of the combined scanner, patients would have a CT exam on one device and PET test on another. Figure 3 shows the design of the hybrid scanner:

 

Figure 3; Design of the scanner (6)

 

Firstly, the patient is injected with a suitable dose of tracer solution depending on the extent of the scan. This is followed by monitoring in a quiet environment at rest.

The patient is positioned as illustrated in figure 3. Initially, a topogram is applied to calculate the scan range. The topogram is a rapid, low dose, planar x-ray exam which identifies anatomical landmarks for positioning of the scan template (7). The template then can be easily resized if necessary. The system is then calibrated for the scan parameters (the patient is scanned from the ear to the mid thigh for a whole body scan (10)) and the spiral CT begins (7).  The main advantage of spiral CT is its ability to scan larger volumes with shorter time periods. This is achieved by connecting the tube voltage cables through a sliding contact positioned on the rotating gantry. The x-ray tubes rotate and the patient table moves in a continuous motion. This technique reduces patient motion, an important factor in the hospital environment. Furthermore, spiral methods give a quicker response to contrast media and reduce motion artifacts (interferences from other planes of the patient) (8). Moreover, the continuity of data along the axis of the patient aids the quality of three-dimensional construction (9).  . The bed is then moved into position and the emission scan begins. On average, acquisition times will range from 4-6 minutes per bed position (10). Ideally the patient should have their arms up. This minimizes beam hardening and artifact effects. Beam hardening is the removal of low energy photons, which reduces contrast. Given the relatively short scanning time, the subject should be able to maintain this position for the duration of the acquisition Patients are advised to maintain constant, shallow breathing during the acquisition [24]. Respiratory motion often results in mushroom artifacts above the diaphragm that are a persistent problem in CT imaging. These problems can be further reduced by instructing the patients to hold their breath at appropriate times during the scan.   The CT dataset provides high quality, anatomical images which are used for PET attenuation correction. This means the scan duration of the PET/CT system is shorter then that from a single PET exam.

 

CT protocols are regularly designed for individual studies on various body parts. The combination of different CT studies in a single examination requires some time and ingenuity to design a procedure that maintains contrast and minimizes the acquisition time and dose to patient. This problem is less prominent in independent imaging centers where information can be analyzed by a single physician.

 

 

Advantages and Limitations of the Scanner

 

There is significant evidence that PET and CT, when analyzed together, increases the detection rate of cancers. Before the introduction of the dual scanner, an abnormality seen on PET could not be corroborated by CT because either there was no obvious lesion or the discontinuity developed in the interim period between scans. As a result, the interpreting physician would have to make an educated guess as to the exact site based exclusively on the anatomical information from the PET data.  Chin et al. (12) found that by sampling 30 patients with mediastinal lymph node involvement the diagnostic accuracy increased to 90%.  Furthermore Vandteenkiste et al. (13) compared FDG-PET and CT images with just CT using lung cancer patients. It was found the diagnostic accuracy increased from 59% to 87%. They also found specificity, accuracy and sensitivity increased from 63%, 68% and 75% to 94%, 95% and 93% respectively. Based on the findings of significant negative predicted value (false positives) using PET alone, it was discovered that mediastinoscopy was unnecessary in 29 of 68 patients (13).

 

However examination of PET and CT images by hand is laborious and vulnerable to human error (10). Before the introduction of the dual scanner, an abnormality seen on PET could not be corroborated by CT because either there was no obvious lesion or the discontinuity developed in the interim period between scans. As a result, the interpreting physician would have to make an educated guess as to the exact site based exclusively on the anatomical information from the PET data. Martinelli et al. (14) tested more then 100 oncology students using a prototype PET/CT scanner. They found that it yielded a more accurate image of FDG uptake, was able to differentiate between physiologic and pathological uptake and led to improvements in monitoring technique. Moreover the same research group (15) discovered it led to significant improvements in patient administration, due to the time saved in interpretation. Furthermore, PET/CT has proven particularly useful in imaging ‘’akward’’ regions of the body such as the post-surgical abdomen and the head/neck. Kamel and co-workers (16) discovered visual FDG uptake in the lower anterior neck in 6 of 184 patients who took part in lung cancer staging. They also found this hybrid devise avoided the false positive effects of PET unaided.

 

The possible advantages of the dual scanner to evaluate patients with lung cancer were investigated by Keidar et al (17). 26 sufferers were sampled which led to improved information regarding lesion localization and FDG uptake in 56% of the patients.

 

A promising clinical role of PET/CT is in radiation treatment planning. CT alone is relatively inaccurate in displaying the size of the tumour. Dizendorf et al. (18) investigated the influence of the integrated device on 30 patients receiving external beam radiation. In 30% of the patients, the dosage was altered and changes in volume and target were recorded in 40% of the patients.

 

Limitations of the Positron Emission Tomography/Computed Tomagraphy Scanner

 

One of the main technical limitations of PET/CT is the problem caused by artifacts due to the breathing of the patient and metallic implants. These difficulties were examined by Osman et al. (19). They concluded incorrect positioning induced by patient movements resulted in 6 of 285 of those tested. They found errors occurred in liver lesions near the base of the right lung.  The group also conducted research on the effect of metallic dental implants on the accuracy of the images. However they found artifacts on the PET images were corrected for by the CT scan (19).

 

The dual scanner signifies a new generation of medical imaging acquisition tools that emphasize the intricacy of multidisciplinary clinical decision making. Perhaps the most demanding area of CT/PET is the necessity to easily display anatomical and functional data that essentially represents 5 or 6 dimensions [24]. The failure to efficiently interpret and navigate through information recorded by the dual scanner is a considerable handicap to many physicians. Currently, no software package exists that provides a user friendly interface to navigate though such detailed, multidimensional data.

 

Figure 4:  Examples of commercially available graphic user interfaces of multimodality viewing software programs [24] A: Siemans B: Hermes C: General Electronic D: Philips  [24]

 

Consequently, image analysis and interpretation can take from 10 minutes in a standard study to 30 minutes in a complicated case with many abnormalities and lesions. This far outweighs the time a physician needs to deduce any irregularities from a standard CT scan. It seems interpretation time is improved in comparison to separate scans but is significantly greater then the single CT scan.

 

A further problem with multimodality diagnostic workstations is their capacity in storing very large data sets. Some scanners will lessen processing time by compromising the spatial resolution from 512*512 pixels per axial slice to 256*256 or by decreasing the range of pixel values from 16 to 8 bits. This compromises the value of the CT image. Consequently considerable processing power is required to maintain the integrity of the image. Even then the loading of a complete, whole body scan can take several minutes.

 

The implementation of image archiving and filmless radiology has improved the value of the PET/CT scanner significantly in recent years. However in large institutions the logistics of sustaining an efficient collaboration between those administering and interpreting the combined data requires flawless communication.  

 

Conclusion and Future Prospects

 

PET/CT is emerging as a preferred method for assessing both anatomy and function of common cancers (10).This is despite drawbacks due to the finite time of the PET scan and the inefficiency of the display mechanisms. However, increased detection and localization accuracy and reduced operating cost mean this hybrid device is fast overtaking separate imagers as the tool of choice in modern hospitals (5).

 

As the usage increases, many of the limitations discussed will be addressed. In particular, appropriate protocols for different cancer types must be formulated and the application of the devise must extend more thoroughly to neurology and cardiology. In addition, the role of the hybrid scanner with other, more disease specific tracers then FDG will become more prominent. Furthermore, it is essential that designers of computer tools and specialists in image communication improve their support of multimodality imaging. The success of PET/CT depends not only on diagnostic improvements but on its application in clinical decision making and patient care.

 

 

References

(1)    http://www.brighamandwomens.org/referringphysiciansnews/PET_CT.pdf

      (2) Medical image analysis. Atam.P.Dhwan, Pg.62

      (3) Medical image analysis. Atam.P.Dhwan, Pg.97

(4)IEEE trans. Med imaging vol 15, 278-289, 2000. S.L Hartmann and R.L Galloway. Depth Buffer targeting for spatial accurate 3-D visualization and medical imaging.

(5)European school of medical physics 2003. Advances in PET (and SPECT) and some applications. Nuclear Medicine image registration. Andrew Todd-Pokropek, UCL.

(6) http://gamma.wustl.edu/rsna03/02-Townsend-RSNA03.pdf

(7) http://www.cpspet.com/our_tech/how_petct_works2.shtml

(8)Medical Imaging physics. 4th edition. William Henden, E. Russel Rilenour, pg 62.

(9)Hu, H.Multislice helical CT: Scan and reconstruction. Med Phys 1999. 26.5.

(10) PET/CT applications and pitfalls. Richard.L.Wahl.M.D. Division of nucleur medicine. John Hopkins Medical Institute, Baltimore, MD.

(11) Dual modality PET/CT ‘’an imaging technology that changes the care of cancer patients’’. John Czernin. Department of molecular and medical pharmacology. Ahmanson biological clinic. UCLA school of medicine. Los Angles, CA.

(12)Chin R Jr, Ward R, Keyes JW, et al. Mediastinal staging of non-small-cell lung cancer with positron emission tomography. Am J Respir Crit Care Med. 1995;152 (6 pt1) 2090-2096.

(13)Vansteenkiste et all. Lymph node staging in non-small-cell lung cancer with FDG-PET scan. A prospective study on 690 lymph node stations from 68 patients. J Clin Oncol. 1998;16:2142-2149.

(14)Martinelli M et al. Survey of results of whole body imaging using PET/CT at the university of Pittsburgh medical centre facility.

(15) Kluetz et al. Combined PET/CT imaging in oncology: Impact on patient management. Clin. Positron Imaging. 2002;85:53-58.

(16)Kamel E, Recurrent laryngeal nerve palsy in patients with lung cancer: detection with PET/CT image fusion. Report of 6 cases. Radiology 2002;224:153-156.

(17)Keidar Z et al. Hybrid imaging using PET/CT with F-18-FDG in suspected recurrence of lung cancer: Diagnostic value and impact on patient management. [abstract] J Nucl Med 2002;43. A-144.

(18)Dizendorf et al, Impact of integrated PET/Ct scanning ion external beam radiation treatment planning [abstract] J Nucl Med. 2002;43:A-547.

(19)Osman M et al. Clinically significant inaccurate localization of lesions with PET-Ct: Frequency in 275 patients [abstract]. J Nucl Med. 2002;43 (suppl): A-116.

[20]Paul Suetens, Fundamentals of Medical Imaging, pg 66 (X-ray Computed Tomography). 

      [21] Ketcham, R.A. and Carlson, W.D., 2001. Acquisition, optimization and interpretation of X-ray computed tomographic imagery: Applications to the geosciences. Computers and Geosciences, 27, 381-400

[22] C390(UCL, Medical Physics Department, Imaging with Ionising Radiation-lecture notes-Robert Speller-Latest Revision June 1998.

[23] Paul Suetens, Fundamentals of Medical Imaging, pg 81 (X-ray Computed Tomography).

[24]Osmin Ratib, MD, PhD:  PET/CT  Image Navigation and Communication, The Journal of Nuclear Medicine vol.45 no.1 Jan 2004. 

[25]Karen M.Mudry, Robert Plonsey, Joseph D.Bronzino, Biomedical Imaging, sect.14-7(PET).

[26] C390(UCL, Medical Physics Department, Imaging with Ionising Radiation-lecture notes- Ian Cullum- sciltillation detector..

[27] http://gsm.utmck.edu/CITDP/documents/principles_000.pdf